Connection and controls are designed for easy operation. |
 |
Patient can fill cylinders while receiving oxygen from the concentrator. |
 |
Gives ambulatory patients greater freedom and independence. |
 |
Virtually eliminates the high service costs of frequent deliveries of cylinders and-or liquid oxygen. |
 |
Small and lightweight. |
 |
Electrical:
120 VAC +- 10% at 60 Hz |
 |
Power: Consumption: 200 watts |
 |
Rated Current Input: 2.0 amps |
 |
Pressure: 14 - 21 psi |
 |
Oxygen: Output from concentrator: 0 - 3 lpm (while filling cylinder) Input: 0 - 2 lpm |
 |
Dimensions: 20.25"W x 15"H x 16"D |
 |
Product Weight: 33 lb |
 |
Cylinder Filling Times: M2: 24 min; ML4: 55 min; ML6: 75 min; M6: 60 min; M9: 125 min; D: 3.5 hrs. |
 |
Quantity: 1 per case |
 |
Warranty: 1 year (3-yr (or 5000-hr) parts & labor on internal-wear components & control-panel components.) |
 |
 |
 |
| HomeFill Ready Rack |
Patient Convenience Pack Bags |
Oxygen cylinders |
| The IOH260 Ready-Rack is designed to fit the Platinum XL and Platinum 10 concentrators. It supports the HomeFill compressor on top of the concentrator, saving space and promoting ease of use. |
A custom bag made just for the Patient Convenience ML4, ML6 or M9 cylinder |
available in sizes, ML4, M6, ML6 and M9. |